TM-XROP-01XN Rev C www.programmablepower.com XHR 1000 Watt Series Programmable DC Power Supply Operation Manual XHR 7.5-130 XHR 20-50 XHR 33-33 X
viii Operating Manual for XHR Series Power SupplyLocal and Remote Sensing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ixList of FiguresFigure 1.1 Power Supply Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Figure 1.2 Rear Panel with Bus
List of Figuresx Operating Manual for XHR Series Power Supply
Operating Manual for XHR Series Power Supply xi List of TablesTable 1.1 SW1 Switch Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
List of Tablesxii Operating Manual for XHR Series Power Supply
15Section 1. Features and SpecificationsDescriptionThe 1000 watt DC output power supplies provide low noise, precisely regulated, variable DC output.
Features and SpecificationsFront Panel Controls16 Operating Manual for XHR Series Power Supply• Remote analog voltage and current limit programming w
Features and SpecificationsFront Panel Controls17Figure 1.1 Power Supply Front Panel(For all models.)AC Power SwitchOutput VoltageControl KnobAC Fail
Features and SpecificationsRear Panel Connectors and Switch18 Operating Manual for XHR Series Power SupplyRear Panel Connectors and SwitchThe output
Features and SpecificationsRear Panel Connectors and Switch19Figure 1.3 Rear Panel with High Voltage Output Connector(For 60 V to 600 V models.)SW1Pro
Features and SpecificationsRear Panel Connectors and Switch20 Operating Manual for XHR Series Power SupplySee Table 1.1 for a list of the functions a
Features and SpecificationsRear Panel Connectors and Switch21See Figure 1.5 to identify the function of each of the 12 terminals on the J2 connector.F
Features and SpecificationsRear Panel Connectors and Switch22 Operating Manual for XHR Series Power SupplyWiringFor lowest noise performance, make co
Features and SpecificationsSpecifications23Specifications ElectricalSpecificationsSpecifications are warranted over a temperature range of 0 to 40 °C
Features and SpecificationsSpecifications24 Operating Manual for XHR Series Power SupplyTable 1.2 Electrical Specifications for 7.5 V to 40 V ModelsM
Features and SpecificationsSpecifications25Table 1.3 Electrical Specifications for 60 V to 600 V ModelsModels 60-18 100-10 150-7 300-3.5 600-1.7O
Features and SpecificationsAdditional Electrical Specifications26 Operating Manual for XHR Series Power SupplyAdditional Electrical SpecificationsInp
Features and SpecificationsElectrical Characteristics27Electrical CharacteristicsRemote Programming and MonitoringSwitching Frequency 7.5 V to 300 V m
Features and SpecificationsEnvironmental Specifications28 Operating Manual for XHR Series Power SupplyEnvironmental SpecificationsApprovalsUnits mark
Features and SpecificationsMechanical Specifications29Mechanical SpecificationsFront Panel Voltage and Current Control10-turn voltage and current pote
About AMETEK AMETEK Programmable Power, Inc., a Division of AMETEK, Inc., is a global leader in the design and manufacture of precision, programmable
Features and SpecificationsDimensions30 Operating Manual for XHR Series Power SupplyDimensionsFigure 1.6 Dimensional Drawings(Dimensions given are no
31Section 2. InstallationBasic Setup ProcedureSee Table 2.1 for a summary of the basic setup procedure and an overall view of the information in this
InstallationInspection, Cleaning, and Packing32 Operating Manual for XHR Series Power SupplyInspection, Cleaning, and PackingInitialInspectionWhen yo
InstallationReturning Power Supplies to the Manufacturer33Returning Power Supplies to the ManufacturerReturnMaterialAuthorizationPolicyBefore returnin
InstallationReturning Power Supplies to the Manufacturer34 Operating Manual for XHR Series Power SupplyPackaging forShipping orStorageFollow these in
InstallationLocation, Mounting, and Ventilation35Location, Mounting, and VentilationRackMountingThe rack mount kit allows you to mount two units side
InstallationAC Input Power36 Operating Manual for XHR Series Power SupplyAC Input PowerWe warrant that the unit will perform to specification at AC i
InstallationFunctional Tests37Functional TestsThe functional test procedures include power-on and front panel function checks as well as voltage and c
InstallationFunctional Tests38 Operating Manual for XHR Series Power SupplyCurrent ModeOperationCheck1. Ensure that the front panel power switch is s
InstallationLoad Connection39Load ConnectionThis section provides recommendations for load wires and how to connect them for both single and multiple
This page intentionally left blank. ii
InstallationLoad Connection40 Operating Manual for XHR Series Power SupplyLoad Wiring Length for Operation with Sense Lines For applications using re
InstallationLoad Connection41Making LoadConnectionsMake load connections to the power supply at output connectors at the rear of the power supply, or
InstallationLoad Connection42 Operating Manual for XHR Series Power SupplyFigure 2.3 Bus Bar ShieldInductiveLoadsUse the following protection methods
InstallationLoad Connection43Figure 2.4 Typical Load Connection Hardware(For 7.5 V to 40 V models.)To make load connections to a 7.5 V, 130 A power su
InstallationLoad Connection44 Operating Manual for XHR Series Power Supply60 V to 600 V ModelsThe output connector for the 60 V to 600 V models is a
InstallationLoad Connection45ConnectingSingle LoadsFigure 2.6 and Figure 2.7 show recommended load and sensing connections for single loads. Local sen
InstallationLoad Connection46 Operating Manual for XHR Series Power SupplyRadial Power Distribution Method To connect distributed loads, we recommend
InstallationLoad Connection47Figure 2.9 Multiple Loads with Remote Sensing
InstallationLocal and Remote Sensing48 Operating Manual for XHR Series Power SupplyLocal and Remote SensingUse connections on the rear panel J2 conne
InstallationLocal and Remote Sensing49Use remote sensing during voltage mode operation to shift the power supply’s regulation point from its output te
Important Safety Instructions Before applying power to the system, verify that your product is configured properly for your particular application.
InstallationLocal and Remote Sensing50 Operating Manual for XHR Series Power Supply.Figure 2.10Connecting Remote Sense Lines
51Section 3. Local OperationStandard OperationOnce you have installed the power supply as covered in Section 2. Installation, the power supply is re
Local OperationStandard Operation52 Operating Manual for XHR Series Power SupplyFigure 3.1 Operating ModesConstant Voltage Mode Operation The power s
Local OperationStandard Operation53ShippedConfiguration(Local ControlMode)The factory ships units already configured for local control (front panel) o
Local OperationUsing Multiple Supplies54 Operating Manual for XHR Series Power SupplyUsing Multiple SuppliesYou can operate power supplies of the sam
Local OperationUsing Multiple Supplies55ConfiguringMultipleSupplies forSeriesOperationUse series operation to obtain a single higher voltage output us
Local OperationUsing Multiple Supplies56 Operating Manual for XHR Series Power SupplyConfiguringMultipleSupplies forParallelOperationUse parallel ope
Local OperationUsing Multiple Supplies57ConfiguringMultipleSupplies forSplit SupplyOperationSplit supply operation uses two power supplies to obtain t
Local OperationUsing Multiple Supplies58 Operating Manual for XHR Series Power SupplyPositive-negative SupplyTo obtain a positive-negative supply, co
Local OperationUsing Over Voltage Protection (OVP)59Using Over Voltage Protection (OVP)The OVP circuit protects the load in the event of a remote prog
This page intentionally left blank. iv
Local OperationUsing Over Voltage Protection (OVP)60 Operating Manual for XHR Series Power SupplyResetting theOVP CircuitTo reset the OVP circuit aft
Local OperationUsing the Shutdown Function61Using the Shutdown FunctionUse the Shutdown function to disable or enable the supply’s output so that you
Local OperationUsing the Shutdown Function62 Operating Manual for XHR Series Power SupplyFigure 3.6 Shutdown Programming with Active Low SignalFigure
Local OperationOver Temperature Protection (OTP)63Over Temperature Protection (OTP)The over temperature protection circuit shuts down the power supply
Local OperationUser Diagnostics64 Operating Manual for XHR Series Power SupplyUser DiagnosticsIf your power supply is not performing as described in
Local OperationUser Diagnostics65Table 3.4 Quick DiagnosticsSymptom Check Further Checks and CorrectionsNo output and the display is blank.Is input vo
Local OperationUser Diagnostics66 Operating Manual for XHR Series Power SupplySymptom Check Further Checks and CorrectionsOutput voltage fluctuating
67Section 4. Remote OperationRemote Analog Programming of Output Voltage and Current LimitRemote analog programming allows control of the output volta
Remote OperationRemote Analog Programming of Output Voltage and Current Limit68 Operating Manual for XHR Series Power SupplyRemoteAnalogProgrammingPr
Remote OperationRemote Analog Programming of Output Voltage and Current Limit69Table 4.1 Power Supply Settings for Different Programming SourcesOutput
Product Family: XHR 1000 Watt Series Programmable DC Power Supply Warranty Period: Five Years WARRANTY TERMS AMETEK Programmable Power, Inc. (“AMETEK
Remote OperationRemote Analog Programming of Output Voltage and Current Limit70 Operating Manual for XHR Series Power SupplyFigure 4.1 Connecting Pro
Remote OperationRemote Monitoring of Output Voltage and Current71Remote Monitoring of Output Voltage and CurrentReadbackSignalsThe J2 connector on the
Remote OperationUsing the Isolated Programming (ISOL) Interface Option72 Operating Manual for XHR Series Power SupplyFigure 4.4 0-10 Vdc Output Volta
Remote OperationUsing the Isolated Programming (ISOL) Interface Option73ISOL SetupandOperatingInstructions1. Ensure that the rear panel SW1 switch is
Remote OperationUsing the Isolated Programming (ISOL) Interface Option74 Operating Manual for XHR Series Power Supply
75Section 5. CalibrationCalibration SetupServiceEnvironmentandPrecautions• Follow established antistatic procedures.• Work at a bench with adequate ro
CalibrationFront Panel Calibration76 Operating Manual for XHR Series Power SupplyFront Panel CalibrationFront panel calibration includes voltmeter, a
CalibrationFront Panel Calibration773. Press the front panel STANDBY switch to its IN position. Check that the output voltage drops to 0V in less than
CalibrationCalibrating for Programming Accuracy78 Operating Manual for XHR Series Power SupplyCalibrating for Programming AccuracyThe factory calibra
CalibrationCalibrating for Programming Accuracy79VoltageProgrammingCircuitCalibration1. Ensure that the power supply is turned OFF. Disconnect any loa
This page intentionally left blank. vi
CalibrationCalibrating for Readback Accuracy80 Operating Manual for XHR Series Power SupplyCalibrating for Readback AccuracyThe factory calibrates th
CalibrationCalibrating for Readback Accuracy81OutputVoltageMonitorCircuitCalibration1. Ensure that the power supply is turned OFF. Disconnect any load
CalibrationCalibrating for Readback Accuracy82 Operating Manual for XHR Series Power SupplyOutputCurrentMonitorCircuitCalibration1. Ensure that the p
viiContentsAbout This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viiList of Figures
Komentáře k této Příručce