Ametek DCS-E 1kW Series Programming Manual Uživatelský manuál Strana 13

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 32
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 12
SECTION 2
ISOLATED ANALOG
PROGRAMMER OPERATION
2.1 SW1 Switch
The SW1 SETUP switch is accessible from the rear panel of the unit. It provides user selection
of the programming/monitoring ranges and shutdown logic signal level. Setting a switch to the 1
or 0 position changes a range. The factory default settings are shown below.
Refer to Figure 2–1 for a rear panel view of the DLM-E 3kW and 4kW models.
Refer to Figure 2–2 for a rear panel view of the DCS-E 1kW models.
Switch
Position
Function OFF (0) Position ON (1) Position
1
Voltage Programming
Range Select
0-10 VDC
* 0-5 VDC, 4-20mA
2
Current Programming
Range Select
0-10 VDC * 0-5 VDC, 4-20mA
3
OVP Programming
Range Select
0-10 VDC * 0-5 VDC
4
Output Voltage Monitor
Range Select
* 0-5 VDC 0-10 VDC
5
Output Current Monitor
Range Select
* 0-5 VDC 0-10 VDC
6
Remote ON-OFF Level
Select
Active Low Signal * Active High Signal
* Indicates default settings
Table 2–1. SW1 SETUP Switch
M51A Option 2-1
Zobrazit stránku 12
1 2 ... 8 9 10 11 12 13 14 15 16 17 18 ... 31 32

Komentáře k této Příručce

Žádné komentáře