Programming Manual Lx \ Ls Series II
66
OUTPut:TTLTrg[:STATe]
This command enables or disables the AC source Trigger Out1 signal, which is available at a
SMA connector on the rear of the Lx\Ls Series units.
Command Syntax OUTPut:TTLTrg[:STATe]<bool>
Parameters 0|1|OFF|ON
*RST Value OFF
Examples OUTP:TTLT 1 OUTP:TTLT OFF
Query Syntax OUTPut:TTLTrg[:STATe]?
Returned Parameters 0 | 1
Related Commands OUTP:TTLT:SOUR
OUTPut:TTLTrg:SOURce
This command selects the signal source for the Trig Out1 signal as follows:
BOT Beginning of transient output
EOT End of transient output
LIST Specified by the TTLTrg list
When an event becomes true at the selected TTLTrg source, a pulse is sent to the SMA
connector on the rear of the AC source.
Command Syntax OUTPut:TTLTrg:SOURce<source>
Parameters BOT | EOT | LIST
*RST Value BOT
Examples OUTP:TTLT:SOUR LIST
Query Syntax OUTPut:TTLTrg:SOURce?
Returned Parameters <CRD>
Related Commands OUTP:TTLT