57
Parameters slew time, refer to model specifications | MIN | MAX
Unit ms
[SOURce:]VOLTage:SLEW:NEGative
This command sets the slew time of the voltage for negative going transitions. MINimum sets the slew
to the fastest possible time. MAXimum sets the slew to the slowest time.
Command Syntax VOLTage:SLEW:NEGative <NRf+>
Parameters Refer to model specifications | MIN | MAX
Unit ms
Query Syntax VOLTage:SLEW:NEGative?
VOLTage:SLEW:NEGative? MIN
VOLTage:SLEW:NEGative? M
AX
Returned Parameters <NRf+> slew time in ms
[SOURce:]VOLTage:SLEW[:POSitive]
This command sets the slew time of the voltage for positive going transitions. MINimum sets the slew
to the fastest possible time. MAXimum sets the slew to the slowest time.
Command Syntax VOLTage:SLEW <NRf+>
VOLTage:SLEW:POSitive <NR1>
Parameters Slew time, refer to model specifications | MIN | MAX
Unit ms
Query Syntax VOLTage:SLEW?
VOLTage:SLEW? MIN
VOLTage:SLEW? MAX
VOLTage:SLEW:POSitive?
VOLTage:SLEW:POSitive? MIN
VOLGage:SLEW:POSitive? MAX
Returned Parameters <NRf+> slew time in ms
[SOURce:]VOLTage:TLEVel, VTR
Channel-Specific Voltage Command/Query. This command specifies the value of the
programmed voltage level for the TRANsient input when the electronic load is in the CV Mode. When
the Transient Subsystem is on, the electronic load input voltage will switch (under control of the
Transient Subsystem) between the main level and TLEVel at a rate determined by the present value
of SLEW.
In order for the input voltage level to switch, TLEVel must be set to a value greater than the main level.
If TLEVel is set to a valu
e belo
w the main level, no error is generated but switching will not occur until
the main level is subsequently below the value of TLEVel.
Command Syntax VOLTage:TLEVel <NRf+>
VTR <NRf+>
Parameters Voltage transient level value, refer individual model specification | MIN |
MAX.
Query Syntax VOLTage:TLEVel?
VOLTage:TLEVel? MIN VOLTage:TLEVel? MAX
VTR?
VTR? MIN
VTR? MAX
Returned Parameters <NRf+> returns the transient voltage level.
If the electronic load is not in CP Mode, the level will still be set, even if it
is less than the presently programmed input level.