M550129-03 Rev K 3-7
Table 3-2. Remote Power-on Conditions
Condition Default
Voltage
0 Volts (initial from factory power-on voltage); otherwise,
last value saved by SCPI command.
See CAL:INIT:VOLT
Current
0 Amps (initial from factory power-on current); otherwise,
last value saved by SCPI command.
See CAL:INIT:CURR
Soft Voltage Limit
Model maximum voltage *
Soft Current Limit
Model maximum current *
OVP Trip Voltage
Model maximum voltage +10% (initial from factory power-
on OVP); otherwise, last value saved by SCPI command.
See CAL:INIT:VOLT:PROT
Delay 0.5 seconds
Foldback Protection OFF (non-configurable)
Output
ON ** See CAL:MOD:POWERON
Hold OFF
Unmask NONE
Service Request
Capability
OFF